FOB Price
Obtener el precio más reciente1 ~ 19999 / Piece ( Negotiable )
|Minimum Order
Place of Origin:
China
Price for Minimum Order:
Minimum Order Quantity:
2 Piece
Packaging Detail:
Poly Case& and Carton.
Delivery Time:
2 week after receiving the deposit.
Supplying Ability:
100 Piece per Week
Payment Type:
T/T, L/C, D/A, D/P, Western Union, Money Gram, PayPal, Other
1. Application
Monitoring of pH in many different industrial processes to reduce cost and optimize yield, especially in harsh conditions such as municipal wastewater and industrial wastewater.
• Water/wastewater treatment
•Inflow of the plant, water collection well, aeration tank, the effluent, etc.
• Pulp and paper, power plant, mine
•Water treatment, chemical treatment feeding
2. Features and Benefits
• The maintenance of electrode is easy and low cost, the reference system repels dirt and rarely clog.
• No electrolyte refilling is required, it can be used under the condition of the max pressure 6bar.
• Strong anti-interference by sulfur.
• History curve function: the measurement data is collected every 5 minutes, data can be stored for one month.
• Notepad function, provides the record of operating and alarming time.
• When the electrode is calibrated, the mode, time and results are recorded, easy to review and analysis the change.
3. Specifications
(1) Performance characteristics
Measuring Range: 0~*4pH, **0~**0°C
Resolution: 0.*1pH, 0.1°C
Accuracy: ±0.*5pH
Repeatability: 1%
Sensitivity: 0.*2pH
Temperature compensation: **0~**0°C,automatic
Calibration: Factory calibrated, also can be calibrated by customer on site
Display: LCD displays data including date, time, measurement value, trendline, etc.
Data Storage: Automatic data store, power-off data protection
(2) Outputs
Analog Outputs: ***0mA/***0 mA output( maximum load **0 ohms)
Relay Outputs: 2 Relays with programmable hysteresis
Relay Capacity: 3A, **0VAC
Digital Communication: MODBUS RS**5,Profibus DP is optional
(3) Power Supply
AC Power Supply: *0~**5 VAC, *0Hz
DC Power Supply: *4 VDC ±*0%
(4) Physical parameters
Dimension of Transmitter: **6×**6×**8mm(W×H×D) (5.7*5.7*4.3 in.)
Dimension of Sensor: Φ*8×**5mm (1.1*6.5 in.)
Materials: *-Transmitter: PC
*-Sensor: Polyester
Process connection of sensor: NPT 3/4
Weight: *-Transmitter: 1Kg
*-Sensor: 0.5Kg
Length of sensor’s Cable: *0m, Shielded cable (other length available)
(5) Environmental conditions
Ingress protection: *-IP*6 (transmitter)
*-IP*8 (Sensor)
Operating Temperature: *-Transmitter: **0~*0°C
*-Sensor: 0~*0°C
Operating Pressure: ≤6Bar
4. Ordering Information
(1) Transmitter Selection
TPH A B CD
*- A, single or duplex isolated 4 ~ *0mA output:
Code Meaning
1 Single isolated 4 ~ *0mA output (pH value)
2 Duplex isolated 4 ~ *0mA output (pH and temperature measured value)
*- B, Signal Output:
Code Meaning
0 4~*0mA
1 RS **5 (Modbus)
2 Profibus DP
*- CD, PowerVersion:
Code Meaning
AC **0VAC
DC *4 VDC
Example
TPH*0AC, pH analyzer transmitter, output Single isolated 4 ~ *0mA (pH), **0V AC Power Supply
(2) Sensor Selection
TPH – S A CXX
*- A, Usage:
Code Meaning
0 Adapted to sewage
1 Other purposes
*- CXX, Cable Length:
Code Meaning
C*0 *0m cable
CXX XX m cable
Example
TPH-S0C*0, a pH sensor,adapted to sewage, with *0m cable.
(3) Accessories
FA***0: Immersion mounting bracket, stainless steel